Bench Top-Vertical Laminar Flow Cabinets

KM Clean Air bench top vertical laminar flow cabinets guarantee a totally clean working environment providing protection for the product or sample that is being handled whilst also preventing cross contamination between samples.

The Vertical laminar flow cabinet works by drawing air in through the intake filter which is situated in the top front panel of the cabinet, all the air then passes through pressure chamber and down through the Hepa filter into the working area of the cabinet being exhausted through the front opening of the cabinet. An advantage with Vertical flow cabinet compared to horizontal flow is that it generates less turbulence around equipment within the cabinet.

With our range of 8 different laminar flow cabinets there is a size to meet all requirements.

All of our vertical laminar flow cabinets our fitted with AAF international HEPA filters or optional ULPA filters.

Minifil e, Minifil & Mikrofil Laminar Flow Cabinets

Vertical laminar flow cabinet works by drawing air in through the intake filter, which is situated in the top front panel of the cabinet, air then passes down through filter into the working area of the cabinet.

The cabinet has been designed with the aid of pharmacists for all aseptic work undertaken in pharmacies, for example mixing of ointments and eye drops, the cabinet is also suitable for biological handling preventing gross contamination, micro electronic work and camera lens cleaning.

See product range technical details and accessories & options for mote information.

Our products are designed and manufactured in Finland. 

Lamil 7, 10, 13, 16, 19 Laminar Flow Cabinets

Vertical laminar flow cabinet works by drawing air in through the intake filter, which is situated in the top front panel of the cabinet, air then passes down through filter into the working area of the cabinet.

The cabinet has been designed with the aid of pharmacists for all aseptic work undertaken in pharmacies, for example mixing of ointments and eye drops, the cabinet is also suitable for biological handling preventing gross contamination, micro electronic work and camera lens cleaning.

See product range technical details and accessories & options for mote information.

Our products are designed and manufactured in Finland. 

Product range technical details

MINIFIL eMINIFILMIKROFILLAMIL 7LAMIL 10LAMIL 13LAMIL 16LAMIL 19
ConstructionPowder painted steel framePowder painted steel framePowder painted steel framePowder painted steel framePowder painted steel framePowder painted steel framePowder painted steel framePowder painted steel frame
External dimensions WxDxH /mm680x455x960680x400x1150680x598x1158700x600x11581000x598x12631300x598x12631600x598x12631900x598x1263
Working area /mm630x375x615680x375x615630x573x615650x575x720950x573x7201250x573x7201550x573x7201850x573x720
Intake filter EU3 /mm320×560255×480255×480255×540255×640255×940255×1240255×1240
HEPA filter H14 / mm610x305x78610x305x69610x305x69610x610x69915x610x691220x610x691524x610x691830x610x69
Lighting /W1515151818303658
Fan /kw0,180,200,240,240,520,520,481,04
Fan controlon/off7 steps7 steps7 steps7 steps7 steps7 steps7 steps
Weight /kg354050587495125155
Noise level / dB (A)5048505055565756
Air velocity m/s0,45(+-20%)0,45(+-20%)0,45(+-20%)0,45(+-20%)0,45(+-20%)0,45(+-20%)0,45(+-20%)0,45(+-20%)
Electrical socketNoYesYesYesYesYesYesYes
Optional u/v lighting /WNo15151530303030

Laminar Accessories & Options

  • Germicidal u/v lightning
  • Stainless steel table with stand
  • Automatic air flow control
  • Night front cover
  • Gas / Vacuum tap
  • Matt black back and side walls
  • Clear Polycarbonate back wall
  • Custom made sizes

Contact us

We help to find suitable clean air solutions for you.

Andrew Siminson

Email. andrew.siminson (at) vectorteam.fi

Tel. +358 40 543 7062